The FGS-03e is a low noise 3-axis fluxgate sensor. It extends the frequency range for MT measurements down to DC. The sensor core unit is a Bartington Mag-03
For lower frequencies all transfer functions are linear. This sensor is designed for crustal studies where periods of 10,000 seconds and more are needed. Recording time for those periods are three to eight weeks depending on the data quality (magnetic activity and external interferences by man made noise).
The fluxgate is measuring the total magnetic field (B). A induction coil is measuring dB/dt.
A good coincidence can be seen for periods greater than 30s. For higher frequencies (smaller periods) a fluxgate can not be used for MT.
Frequency range | DC ..... 3 kHz DC – 30 sec for MT measurements |
Sensor noise | < 0.006 nT √Hz @ 1Hz |
Output sensitivity | 0.1 mV/nT (model L100) 0.143 mV/nT (model L70) |
Orthogonality error | < 0.5° |
Output voltage range | ± 10V |
Measuring range | ± 100,000 nT (model L100), ± 70,000 nT (model L70) |
Connector | ODU G32KON-T08QJ00-000 (FluxGate socket) ODU S22KON-T08MJG0-7000 (cable plug) |
Offset temperature coefficient | ± 0.1 nT / °C |
Offset error | ± 5 nT |
Supply voltage | ± 12V to ± 15V stabilized and filtered |
Supply current | +26mA, -6mA (+1.4mA per 100μT for each axis) |
Case | ruggedized, waterproof case |
Weight | approx. 3 kg |
External dimensions | length 300 mm, Ø 75mm |
Operating temperature range | -40°C ..... + 70°C |
Description |
---|
Downloads FGS-03e |
This website uses cookies. Those have two functions: On the one hand they are providing basic functionality for this website. On the other hand they allow us to improve our content for you by saving and analyzing anonymized user data. You can redraw your consent to to using these cookies at any time. Find more information regarding cookies on our Data Protection Declaration and regarding us on the Imprint.