The electric field probe EFP-07 is a Ag/AgCl (silver/ silver-chloride) electrode.
The electrode can be filled with environmental friendly saturated NaCl or KCl solution (3M NaCl or 3.5 M Kcl).
This electrolyte can be made with 28g/l NaCl (normal salt) in distilled water (pure H2O).
The electrode needs 8-12 hours of pre-soaking in the electrolyte before being used.
Do not use contaminated water or water enhanced with chemical agents. This may seriously degrade the performance of the half-cell and might lead to a damage of the surface.
Measuring the contact resistivities between two electrodes, e.g. when placed in the ground or in an electrolyte bath, will degrade the electrode (cathode (-) and anode (+)). You will irreversible harm the Ag/AgCl electrode. Do not use these electrodes to inject currents!
Tiny currents may not harm.
| Frequency range | DC .... 50 kHz |
| Contact resistance | 50 – 500 Ω (for normal soils) |
| Effective surface | 2 square centimetre |
| DC potential | 0.1 to 0.9 mV |
| Temperature Range | +5 to +40 °C |
| Temperature Drift | n μV/°C the temperature drift of two electrodes in the current circle is much lower than n μV/°C because of the same electrical behaviour; electrodes can be used at a wider outside temperature range if buried deep enough |
| Weight | Appr. 1.05 kg |
| Working Life | 4-5 years; careful usage |
| External dimensions | length 165 mm, diameter 110 mm |